LY3110H Stationary Single Phase Meter Test Bench

Introduction

Used for testing single-phase energy meters in Measurement Test Centers, Utilies,National Insitute of Metrology, Meter manufactures and for R&D.

The number of test positions is flexible. The photo shows a configuration with 24 positions. 


Deprecated: trim(): Passing null to parameter #1 ($string) of type string is deprecated in /home/u427399520/domains/szlongyuan.com/public_html/wp-content/plugins/tablepress/classes/class-elementor-widget-table.php on line 260
ItemValues
Reference StandardClass 0.05,optional Class 0.1
Voltage range220V
Current range0.1A,0.5A,1A,5A,10A,20 A,50A,100 A
Voltage output capacity500 VA
Current output capacity2000 VA
Voltage-and current adjustmen0%-120%×Range,minimum voltage 30V
Phase shift±180°
Test frequency40 Hz-70 Hz
Power stability≤0.01%/100 s
Distortion≤0.5%
HarmonicsUp to 21
Harmonics amplitude≤40%
Hammonics phase shift0°-360°
Pulse outputMax.50 kHz
Pulse inputMax.10 kHz
PositionsClass 0.05: Max.48,each position with galvanic separated voltage
StructureReference standard and source integrated in rack
Operating power source3×220 V/380V±15%,45 Hz-65 Hz(other voltages or frequencies on request)
Operating temperature23℃±10℃
Humidity≤85%